Κυριακή 18 Δεκεμβρίου 2022

Comparison of Oura Smart Ring Against ActiGraph Accelerometer for Measurement of Physical Activity and Sedentary Time in a Free-Living Context

alexandrossfakianakis shared this article with you from Inoreader
imageSmart rings, such as the Oura ring, might have potential in health monitoring. To be able to identify optimal devices for healthcare settings, validity studies are needed. The aim of this study was to compare the Oura smart ring estimates of steps and sedentary time with data from the ActiGraph accelerometer in a free-living context. A cross-sectional observational study design was used. A convenience sample of healthy adults (n = 42) participated in the study and wore an Oura smart ring and an ActiGraph accelerometer on the non-dominant hand continuously for 1 week. The participants completed a background questionna ire and filled out a daily log about their sleeping times and times when they did not wear the devices. The median age of the participants (n = 42) was 32 years (range, 18–46 years). In total, 191 (61% of the potential) days were compared. The Oura ring overestimated the step counts compared with the ActiGraph. The mean difference was 1416 steps (95% confidence interval, 739–2093 steps). Daily sedentary time was also overestimated by the ring; the mean difference was 17 minutes (95% confidence interval, −2 to 37 minutes). The use of the ring in nursing interventions needs to be considered.
View on Web

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Αρχειοθήκη ιστολογίου